  4. Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling
 
research article

Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling

Wegrzyn, Julien
•
Antoniadis, Alexander
•
Sarshari, Ehsan  
Show more
January 12, 2022
International Orthopaedics

Purpose Concerns remain about potential increased wear with dual mobility cups related to the multiple articulations involved in this specific design of implant. This finite element analysis study aimed to compare polyethylene (PE) wear between dual mobility cup and conventional acetabular component, and between the use of conventional ultra-high molecular weight PE (UHMWPE) and highly cross-linked PE (XPLE).

Methods Patient-specific finite element modeling was developed for 15 patients undergoing primary total hip arthroplasty (THA). Five acetabular components were 3D modeled and compared in THA constructs replicating existing implants: a dual mobility cup with a 22.2-mm-diameter femoral head against UHMWPE or XLPE (DM22PE or DM22XL), a conventional cup with a 22.2-mm-diameter femoral head against UHMWPE (SD22PE) and a conventional cup with a 32-mm-diameter femoral head against UHMWPE or XLPE (SD32PE or SD32XL).

Results DM22PE produced 4.6 times and 5.1 times more volumetric wear than SD32XL and DM22XL (p < 0.0001, Cohen's d = 6.97 and 7.11; respectively). However, even if significant, the differences in volumetric wear between DM22XL and SD32XL as well as between DM22PE and SD22PE or SD32PE were small according to their effect size (p < 0.0001, Cohen's vertical bar d vertical bar= 0.48 to 0.65) and could be therefore considered as clinically negligible.

Conclusion When using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional acetabular component with a 32-mm-diameter femoral head against XLPE. Therefore, XLPE is advocated in dual mobility cup to improve its wear performance.

  • Files
  • Details
  • Metrics
Loading...
Thumbnail Image
Name

s00264-022-05305-4.pdf

Type

Publisher's Version

Version

http://purl.org/coar/version/c_970fb48d4fbd8a85

Access type

openaccess

License Condition

CC BY

Size

1.21 MB

Format

Adobe PDF

Checksum (MD5)

fe623d457c9ea21306d4c156086ed2e3

Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés