Polyethylene wear of dual mobility cups: a comparative analysis based on patient-specific finite element modeling

Wegrzyn, Julien
•
Antoniadis, Alexander
•
Sarshari, Ehsan  
Show more
January 12, 2022
International Orthopaedics

Purpose Concerns remain about potential increased wear with dual mobility cups related to the multiple articulations involved in this specific design of implant. This finite element analysis study aimed to compare polyethylene (PE) wear between dual mobility cup and conventional acetabular component, and between the use of conventional ultra-high molecular weight PE (UHMWPE) and highly cross-linked PE (XPLE).

Methods Patient-specific finite element modeling was developed for 15 patients undergoing primary total hip arthroplasty (THA). Five acetabular components were 3D modeled and compared in THA constructs replicating existing implants: a dual mobility cup with a 22.2-mm-diameter femoral head against UHMWPE or XLPE (DM22PE or DM22XL), a conventional cup with a 22.2-mm-diameter femoral head against UHMWPE (SD22PE) and a conventional cup with a 32-mm-diameter femoral head against UHMWPE or XLPE (SD32PE or SD32XL).

Results DM22PE produced 4.6 times and 5.1 times more volumetric wear than SD32XL and DM22XL (p < 0.0001, Cohen's d = 6.97 and 7.11; respectively). However, even if significant, the differences in volumetric wear between DM22XL and SD32XL as well as between DM22PE and SD22PE or SD32PE were small according to their effect size (p < 0.0001, Cohen's vertical bar d vertical bar= 0.48 to 0.65) and could be therefore considered as clinically negligible.

Conclusion When using XLPE instead of UHMWPE, dual mobility cup with a 22.2-mm-diameter femoral head produced a similar amount of volumetric wear than conventional acetabular component with a 32-mm-diameter femoral head against XLPE. Therefore, XLPE is advocated in dual mobility cup to improve its wear performance.

  • Files
  • Details
  • Metrics
Type
research article
DOI
10.1007/s00264-022-05305-4
Web of Science ID

WOS:000741926700004

Author(s)
Wegrzyn, Julien
Antoniadis, Alexander
Sarshari, Ehsan  
Boubat, Matthieu  
Terrier, Alexandre  
Date Issued

2022-01-12

Publisher

SPRINGER

Published in
International Orthopaedics
Volume

46

Start page

779

End page

787

Subjects

Orthopedics

•

dual mobility cup

•

polyethylene mobile component

•

patient-specific modeling

•

polyethylene wear

•

highly cross-linked polyethylene

•

total hip-arthroplasty

•

dislocation

•

liners

•

simulator

•

bearings

•

friction

Editorial or Peer reviewed

REVIEWED

Written at

EPFL

EPFL units
LBO  
Available on Infoscience
January 31, 2022
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/184917
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés