research article
Miniature Electrochemical Glucose Biosensors
Glucose-oxidase-based sensors were realized using three techniques of immobilizing thin layers of enzyme over a three-thin-film electrode transducer. Sensors with chemically crosslinked membranes, completed by an outer diffusion-limiting membrane, were characterized with consideration of the basic requirements of an implantable device. Thus, e.g. a negligible loss of the activity was observed when the sensors were sterilized by a 2.5 MRad dose of gamma radiation. On the other hand, the sensors with the electrochemically deposited membranes were particularly suitable, e.g. with respect to response time, etc., for flow-injection analysis measurements.
Type
research article
Author(s)
Date Issued
1993
Published in
Volume
281
Issue
3
Start page
461
End page
466
Note
98
Editorial or Peer reviewed
REVIEWED
Written at
OTHER
EPFL units
Available on Infoscience
May 12, 2009
Use this identifier to reference this record