Rapid and precise in vivo measurement of human corneal thickness with optical low-coherence reflectometry in normal human eyes
An optical low-coherence reflectometer is used for rapid noncontact measurements of the human corneal thickness in vivo. Thickness measurements on ten volunteers show a standard deviation of 3.4 μm. The experiments reveal that the optical reflectometer benefits from a 2.5 fold enhancement of the measurement precision and a 2.8 fold reduction in measurement time compared to a standard clinical ultrasonic pachometer
1998
3
3
253
8
Inst. of Appl. Opt., Swiss Fed. Inst. of Technol., Lausanne, Switzerland Copyright 1998, IEE 5974792 1083-3668 rapid noncontact measurements human corneal thickness volunteers optical reflectometer measurement precision enhancement standard clinical ultrasonic pachometer measurement time reduction normal human eyes optical low-coherence reflectometry rapid precise in vivo measurement