  4. in vivo Electrical Impedance Spectroscopy of Tissue Reaction to Microelectrode Arrays
 
research article

in vivo Electrical Impedance Spectroscopy of Tissue Reaction to Microelectrode Arrays

Mercanzini, André  
•
Colin, P.  
•
Bensadoun, J.-C.  
Show more
2009
IEEE Transactions on Biomedical Engineering

The goal of this experiment was to determine the electrical properties of the tissue reaction to implanted microelectrode arrays. We describe a new method of analyzing electrical impedance spectroscopy data to determine the complex impedance of the tissue reaction as a function of postimplantation time. A model is used to extract electrical model parameters of the electrode-tissue interface, and is used to isolate the impedance of the tissue immediately surrounding the microelectrode. The microelectrode arrays consist of microfabricated polyimide probes, incorporating four 50-mum- diameter platinum microelectrodes. The devices were implanted in the primary motor cortex of adult rats, and measurements were performed for 12 weeks. Histology was performed on implants at three time points in one month. Results demonstrate that the tissue reaction causes a rapid increase in bioimpedance over the first 20 days, and then stabilizes. This result is supported by histological data.

  • Files
  • Details
  • Metrics
Type
research article
DOI
10.1109/TBME.2009.2018457
Web of Science ID

WOS:000267379100015

Author(s)
Mercanzini, André  
Colin, P.  
Bensadoun, J.-C.  
Bertsch, Arnaud  
Renaud, Philippe  
Date Issued

2009

Published in
IEEE Transactions on Biomedical Engineering
Volume

56

Issue

7

Start page

1909

End page

1918

Subjects

Biomedical microdevices

•

impedance spectroscopy

•

microelectrode arrays

•

neuroprostheses

Editorial or Peer reviewed

REVIEWED

Written at

EPFL

EPFL units
LMIS4  
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/42504
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés