Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry
The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.
2-s2.0-85218626437
Istituto Nazionale di Fisica Nucleare, Sezione di Torino
Istituto Nazionale di Fisica Nucleare, Sezione di Torino
Istituto Nazionale di Fisica Nucleare, Sezione di Torino
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
2025-01-01
20
1
C01034
REVIEWED
EPFL