Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye
A parallel Frequency Domain Optical Coherence Tomography (FD-OCT) system and - to the best of our knowledge- first in vivo tomograms obtained with such system are presented. A full tomogram of 256(x) x 512(z) pixels covering a sample region of 8 mm x 3,8 mm is recorded in only 1 ms. Since the transverse as well as the depth information is obtained in parallel, the structure is free of any motion artifacts. In order to study cross talk issues for parallel illumination the transversal resolution for a thermal light source is compared to that with an SLD.
oe-13-4-1131.pdf
Publisher's Version
http://purl.org/coar/version/c_970fb48d4fbd8a85
openaccess
n/a
631.39 KB
Adobe PDF
49ecb948ee3c1fb0acc215560a727ddd