Design and Modelling of a Super-Regenerative Receiver for Medical Implant Devices

Medical implant devices have been widely used in recent years. The Super-Regenerative Receiver has been on preferred architecture due to its power advantage over other architectures. We present a detailed analysis of the circuits and their equivalent models to be used in system level design of a Super-Regenerative Receiver in this paper. Designs were carried out in UMC 180nm process with a center frequency of 416MHz for MedRadio band. The study is concluded with the simulation results of the circuits and the equivalent models.


Published in:
2018 14Th Conference On Phd Research In Microelectronics And Electronics (Prime 2018), 245-248
Presented at:
15th International Conference on Synthesis, Modeling, Analysis and Simulation Methods and Applications to Circuit Design (SMACD) / 14th Conference on PhD Research in Microelectronics and Electronics (PRIME), Prague, CZECH REPUBLIC, Jul 02-05, 2018
Year:
Jan 01 2018
Publisher:
New York, IEEE
ISBN:
978-1-5386-5386-9
Laboratories:




 Record created 2019-11-26, last modified 2020-04-20


Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)