An Integrated Platform for Differential Electrochemical and ISFET Sensing
A fully-integrated differential biosensing platform on CMOS is presented for miniaturized enzyme-based electrochemical sensing. It enables sensor background current elimination and consists of a differential sensor array and a <i>differential readout IC</i> (DiRIC). The sensor array includes a four-electrode sensor for amperometric electrochemical sensing, as well as a differential ISFET-based pH sensor to calibrate the biosensors. The ISFET is biased in weak inversion and co-designed with DiRIC to enable pH measurements from 1 to 14 with resolution of 0.1 pH. DiRIC enables differential current measurement in the range of ±100 μA with more than 120dB dynamic range.
Sara-Pantelis-ISCAS16-Differenzial.pdf
openaccess
772.25 KB
Adobe PDF
40748f068d1988a8884e4d81fd6befa1