2014

This document concerns the deviation of the output of MilliNewton and CentoNewton force sensors from ideal ratiometric behaviour. These sensors nominally operate from a single +5 V supply voltage and, as they have no voltage regulator, their output response should be ideally "ratiometric", i.e. proportional to supply voltage, within the operating range of the used amplifier chip. However, due to the amplifier characteristics (nonlinear offset), a substantial deviation of the sensor offset (zero-load output) from this ideal behaviour can appear if the supply voltage strongly deviates from 5 V. On the other hand, the span remains essentially ratiometric, and sensor functionality is maintained.

  • Files
  • Details
  • Metrics
Loading...
Thumbnail Image
Name

T0003-EN mN ratiometric EPFL - 2014-07-16.pdf

Access type

openaccess

Size

943.85 KB

Format

Adobe PDF

Checksum (MD5)

f22ecc5dffad2eb4603865ecd811cacd

Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés