Abstract

Theoretically the auscultatory method using Korotkoff sounds is more related to the maximum artery closure status, while the oscillometric method is more related to the overall artery closure status under the cuff. Therefore, the latter is less accurate than the former. This work introduces a new method, which is more accurate than the oscillometric method and suitable for automatic devices. To monitor the maximum artery closure status, a piezoelectric film sensor is attached to the skin just above the brachial artery and under the central section of the cuff where maximum cuff pressure is transferred to the arm. Using the waveform features obtained by this sensor, measurement errors of 0.7 +/- 2.5 and 1.27 +/- 4.53 mmHg were obtained for the systolic and diastolic pressure, respectively. These reflect small deviations from auscultatory clinical data. (C) 2012 Elsevier Ltd. All rights reserved.