Dual-Channel Electrospray Microchip

A dual-channel electrospray microchip has been developed for electrospray ionization mass spectrometry (ESI-MS) where aqueous samples are mixed at the Taylor cone with an organic buffer. Due to the fast and effective mixing in the Taylor cone, the aqueous sample can be well ionized with a high ion intensity. The influence of geometric parameters such as the distance between the two microchannels at their junction at the tip of the emitter has been investigated together with chemical parameters such as the organic buffer composition.


Published in:
Journal of The American Society for Mass Spectrometry, 24, 3, 454-457
Year:
2013
ISSN:
1879-1123
Laboratories:




 Record created 2013-03-19, last modified 2018-03-17

Publisher's version:
Download fulltextPDF
Supporting Information:
Download fulltextDOC
TOC graphic:
Download fulltextGIF
Rate this document:

Rate this document:
1
2