Electrostatic-Spray Ionization Mass Spectrometry

An electrostatic-spray ionization (ESTASI) method has been used for mass spectrometry (MS) analysis of samples deposited in or on an insulating substrate. The ionization is induced by a capacitive coupling between an electrode and the sample. In practice, a metallic electrode is placed close to but not in direct contact with the sample. Upon application of a high voltage pulse to the electrode, an electrostatic charging of the sample occurs leading to a bipolar spray pulse. When the voltage is positive, the bipolar spray pulse consists first of cations and then of anions. This method has been applied to a wide range of geometries to emit ions from samples in a silica capillary, in a disposable pipet tip, in a polymer microchannel, or from samples deposited as droplets on a polymer plate. Fractions from capillary electrophoresis were collected on a polymer plate for ESTASI MS analysis.


Published in:
Analytical Chemistry, 84, 17, 7422-7430
Year:
2012
Publisher:
Washington, Amer Chemical Soc
ISSN:
1520-6882
Laboratories:




 Record created 2012-10-11, last modified 2018-03-17

Publisher's version:
Download fulltextPDF
Supporting Information:
Download fulltextPDF
TOC graphic:
Download fulltextGIF
Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)