Abstract
In this contribution we present our work on filtration of plasma from whole blood in a polymer-based microfluidic system. The absorption of pathogens was found to be low; therefore our system is valid for clinical trials. The blood filtration element is capable of extracting 8–15μL volume of plasma from 100μL of whole blood in 7 minutes.
Details
Title
Clinical Validation of a Blood Filtration Element for Plasma Separation at the Point of Care
Author(s)
Homsy, Alexandra ; van der Wal, Peter ; Doll, Werner ; Schaller, Roland ; Korsatko, Stefan ; Ratzer, Maria ; Ellmerer, Martin ; Pieber, Thomas ; Nicol, Andreas ; de Rooij, Nico