  4. Reference-Free Automated Magnetic Sensor Calibration for Angle Estimation in Smart Knee Prostheses
 
research article

Reference-Free Automated Magnetic Sensor Calibration for Angle Estimation in Smart Knee Prostheses

Arami, Arash  
•
Rechenmann, Julien  
•
Aminian, Kamiar  
2014
IEEE Sensors Journal

In this work, we present a method for automated calibration of an implanted anisotropic magnetoresistive (AMR) sensor for measuring the internal-external rotation (IE) in a prosthetic knee without using any reference measurement. The measurement system consists of a permanent magnet and a 2-D AMR sensor configured and embedded into the prosthesis. Using the physical relation between the measurements of channels of sensor, we proposed a reference-free calibration method through the optimization. The parameters of the measurement model of the sensor obtained via the calibration were directly used in an angle estimator. Two different optimization approaches were tested, and the resultant estimates were compared to the reference angle with the expected ± standard deviation of error of -1.69°±0.65° and -1.17°±1.13°. The reference-free calibrated estimators were also compared to the reference-dependent least-square estimators. This latter comparison showed an offset error difference among the estimators while the coefficients of determination for the both reference-free and reference-dependent estimators were high, i.e. larger than 0.91. The good performances were maintained even when using a small portion of data for the optimization. The sensor-magnet misalignment effect was also investigated. The misalignment slightly worsened the mean error of the estimates, but neither dramatically affected the variance of error nor the coefficient of determination.

  • Details
  • Metrics
Type
research article
DOI
10.1109/JSEN.2014.2303644
Web of Science ID

WOS:000334685600001

Author(s)
Arami, Arash  
Rechenmann, Julien  
Aminian, Kamiar  
Date Issued

2014

Publisher

Institute of Electrical and Electronics Engineers

Published in
IEEE Sensors Journal
Volume

14

Issue

6

Start page

1788

End page

1796

Subjects

Knee rotations

•

Smart knee prosthesis

•

Anisotropic Magnetoresistive sensor

•

Optimization

URL

URL

http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=6728606
Editorial or Peer reviewed

REVIEWED

Written at

EPFL

EPFL units
LMAM  
Available on Infoscience
January 17, 2014
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/99580
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés