  4. Fabrication of multilayer systems combining microfluidic and micro-optical elements for fluorescence detection
 
research article

Fabrication of multilayer systems combining microfluidic and micro-optical elements for fluorescence detection

Roulet, J.-C.
•
Völkel, R.
•
Herzig, H. P.  
Show more
2001
Journal of Microelectromechanical Systems

This paper presents the fabrication of a microchemical chip for the detection of fluorescence species in microfluidics. The microfluidic network is wet-etched in a Borofloat 33 (Pyrex) glass wafer and sealed by means of a second wafer. Unlike other similar chemical systems, the detection system is realized with the help of microfabrication techniques and directly deposited on both sides of the microchemical chip. The detection system is composed of the combination of refractive microlens arrays and chromium aperture arrays. The microfluidic channels are 60 μm wide and 25 μm deep. The utilization of elliptical microlens arrays to reduce aberration effects and the integration of an intermediate (between the two bonded wafers) aluminum aperture array are also presented. The elliptical microlenses have a major axis of 400 μm and a minor axis of 350 μm. The circular microlens diameters range from 280 to 300 μm. The apertures deposited on the outer chip surfaces are etched in a 3000-Å-thick chromium layer, whereas the intermediate aperture layer is etched in a 1000-Å-thick aluminum layer. The overall thickness of this microchemical system is less than 1.6 mm. The wet-etching process and new bonding procedures are discussed. Moreover, we present the successful detection of a 10-nM Cy5 solution with a signal-to-noise ratio (SNR) of 21 dB by means of this system.

  • Details
  • Metrics
Type
research article
DOI
10.1109/84.967369
Author(s)
Roulet, J.-C.
Völkel, R.
Herzig, H. P.  
Verpoorte, E.
de Rooij, N. F.  
Dändliker, R.  
Date Issued

2001

Publisher

Institute of Electrical and Electronics Engineers

Published in
Journal of Microelectromechanical Systems
Volume

10

Issue

4

Start page

482

End page

491

Note

256

Editorial or Peer reviewed

REVIEWED

Written at

OTHER

EPFL units
SAMLAB  
OPT  
Available on Infoscience
May 12, 2009
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/39813
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés