Comparative study of human erythrocytes by digital holographic microscopy, confocal microscopy, and impedance volume analyzer

Rappaz, Benjamin  
•
Barbul, Alexander
•
Emery, Yves
Show more
2008
Cytometry Part A

Red blood cell (RBC) parameters such as morphology, volume, refractive index, and hemoglobin content are of great importance for diagnostic purposes. Existing approaches require complicated calibration procedures and robust cell perturbation. As a result, reference values for normal RBC differ depending on the method used. We present a way for measuring parameters of intact individual RBCs by using digital holographic microscopy (DHM), a new interferometric and label-free technique with nanometric axial sensitivity. The results are compared with values achieved by conventional techniques for RBC of the same donor and previously published figures. A DHM equipped with a laser diode ( = 663 nm) was used to record holograms in an off-axis geometry. Measurements of both RBC refractive indices and volumes were achieved via monitoring the quantitative phase map of RBC by means of a sequential perfusion of two isotonic solutions with different refractive indices obtained by the use of Nycodenz (decoupling procedure). Volume of RBCs labeled by membrane dye Dil was analyzed by confocal microscopy. The mean cell volume (MCV), red blood cell distribution width (RDW), and mean cell hemoglobin concentration (MCHC) were also measured with an impedance volume analyzer. DHM yielded RBC refractive index n = 1.418 ± 0.012, volume 83 ± 14 fl, MCH = 29.9 pg, and MCHC 362 ± 40 g/l. Erythrocyte MCV, MCH, and MCHC achieved by an impedance volume analyzer were 82 fl, 28.6 pg, and 349 g/l, respectively. Confocal microscopy yielded 91 ± 17 fl for RBC volume. In conclusion, DHM in combination with a decoupling procedure allows measuring noninvasively volume, refractive index, and hemoglobin content of single-living RBCs with a high accuracy. © 2008 International Society for Advancement of Cytometry

  • Details
  • Metrics
Type
research article
DOI
10.1002/cyto.a.20605
Web of Science ID

WOS:000259750300005

Author(s)
Rappaz, Benjamin  
Barbul, Alexander
Emery, Yves
Korenstein, Rafi
Depeursinge, Christian  
Magistretti, Pierre J.
Marquet, Pierre
Date Issued

2008

Published in
Cytometry Part A
Volume

73A

Issue

10

Start page

895

End page

903

Subjects

[MVD]

•

digital holographic microscopy

•

laser scanning confocal microscopy

•

impedance volume analyzer

•

red blood cell volume

•

MCV

•

refractive index

•

mean corpuscular hemoglobin concentration

•

MCHC

Editorial or Peer reviewed

REVIEWED

Written at

EPFL

EPFL units
LOA  
Available on Infoscience
July 20, 2009
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/41653
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés