Gini, E.Spiekman, L.H.van Brug, H.Vinchant, J.-F.Morasca, S.Pozzi, F.De Bernardi, C.Zengerle, R.Weiershausen, W.Noell, W.Thévenaz, L.Küng, A.Enard, A.Vodjdani, N.2015-11-112015-11-112015-11-111994https://infoscience.epfl.ch/handle/20.500.14299/120556The results of round-robin measurements of waveguide loss and integrated mirror reflectivity in several European laboratories in the frame of the COST 240 project are presented. The Fabry-Perot cavity contrast method was used on semiconductor optical waveguides. The measured losses show a satisfactory agreement (within 0.5 dB) between all the laboratories, but also point out some aspects which must be taken into account to get meaningful data.Measurements of loss and mirror reflectivity in semiconductor optical waveguides: A European interlaboratory comparison experimenttext::conference output::conference proceedings::conference paper