Cuchet, C.Espanet, Ch.Civet, Y.Biwersi, S.Perriard, Y.2016-10-282016-10-282016-10-28201610.1109/AIM.2016.7576859https://infoscience.epfl.ch/handle/20.500.14299/130823WOS:000387100300123The proposed paper discusses a new methodology to monitor the B-H curve of magnetic fluids, which have low relative permeability. The set-up consists of a magnetic circuit which encloses a sample. A first coil allows creating a magnetic field into the sample whereas a second one measures the flux density. A reluctance network taking into account the leakage flux is investigated in order to improve the result of a simple reluctance network. In fact, a comparison of the two reluctance networks is effectuated according to the Vibrating Sample Magnetometer measurement, which gives an error up to 80% for the simple reluctance network, while the relative error is equal to 20% with the new developed reluctance network presented in this paper.Benchmark testingCoilsDensity measurementMagnetic fluidsMagnetic fluxMagnetic permeability measurementMagnetometersMeasurement errorsVibration measurementLeakage Flux Modelling in Magnetic Fluids Test Benchtext::conference output::conference proceedings::conference paper