Gamby, J.Abid, J.-P.Girault, H. H.2005-11-072005-11-072005-11-07200510.1021/ja050602ohttps://infoscience.epfl.ch/handle/20.500.14299/219409WOS:000232170800061A sensor for measuring adsorption on a substrate has been designed including a contactless detection scheme, called supercapacitive admittance tomoscopy (SCAT). The sensor comprises a thin dielectric layer with two parallel band electrodes on the one side and a chemically modified surface on the other onto which the adsorption of molecules occurs. Upon application of a high frequency ac voltage between the two electrodes, a capacitive coupling is established across the dielectric layer, and the admittance measured depends on the surface state of the chemically modified interface. On the basis of this principle, a flow sensor has been developed to measure sensorgrams to follow the dynamics of the adsorption and has been tested for the adsorption of IgG on the modified surface.Supercapacitive Admittance Tomoscopytext::journal::journal article::research article