Spectroscopic optical coherence tomography based on wavelength de-multiplexing and smart pixel array detection

Current spectroscopic optical coherence tomography (OCT) methods rely on a posteriori numerical calculation. We present an experimental alternative for accessing spectroscopic information in OCT without post-processing based on wavelength de-multiplexing and parallel detection using a diffraction grating and a smart pixel detector array. Both a conventional A-scan with high axial resolution and the spectrally resolved measurement are acquired simultaneously. A proof-of-principle demonstration is given on a dynamically changing absorbing sample. The method's potential for fast spectroscopic OCT imaging is discussed. The spectral measurements obtained with this approach are insensitive to scan non-linearities or sample movements. (C) 2004 Elsevier B.V. All rights reserved.


Published in:
Optics Communications, 237, 4-6, 275-283
Year:
2004
ISSN:
0030-4018
Keywords:
Note:
Swiss Fed Inst Technol, Lab Opt Biomed, CH-1015 Lausanne, Switzerland. Laubscher, M, Swiss Fed Inst Technol, Lab Opt Biomed, CH-1015 Lausanne, Switzerland. markus.laubscher@epfl.ch
ISI Document Delivery No.: 833YD
Times Cited: 1
Cited Reference Count: 17
Other identifiers:
Laboratories:




 Record created 2006-10-03, last modified 2018-03-17