Use of microbalances for measurement of gas sorption on catalysts

The use of microbalances to study gas-solid reactions at high temps. and normal pressure is often limited by fluctuations due to thermal convection in the reaction zone. The fluctuations depend on the pressure, the type of gas, the temp., and the geometry of the arrangement. The problems were overcome by distributing the gas on the circumference of the tube surrounding the specimen. The obsd. fluctuations under normal pressure (1.1 bar) and temps. ?200 Deg were <0.5 mg. [on SciFinder (R)]


Published in:
Chimia, 34, 4, 181-4
Year:
1980
Keywords:
Note:
CAN 92:221481
67-1
Catalysis and Reaction Kinetics
Ec. Polytech. Fed.,Lausanne,Switz.
Journal
0009-4293
written in German.
Laboratories:




 Record created 2006-04-18, last modified 2018-03-17


Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)