Journal article

Femtosecond fluorescence upconversion setup with broadband detection in the ultraviolet

We show a femtosecond fluorescence upconversion setup with broadband detection to measure time-resolved emission spectra in the 300-550 nm range, upon excitation between 250 and 300 nm, with a time resolution of 100 fs. We present time-resolved fluorescence emission spectra of 2,5-diphenyloxazole in solution, which demonstrate the capabilities of the setup.


    Ecole Polytechnique Federale de Lausanne, Laboratoire de Spectroscopie Ultrarapide, Institut des Sciences et Ingenierie Chimiques, CH-1015 Lausanne, Switzerland

    0146-9592 FIELD Electronic Internat.Standard Doc. Number:

    United States

    Journal; Article; (JOURNAL ARTICLE)



    Record created on 2006-02-27, modified on 2016-08-08