Note: Versatile sample stick for neutron scattering experiments in high electric fields

We present a versatile high voltage sample stick that fits into all cryomagnets and standard cryostats at the Swiss Spallation Neutron Source, Paul Scherrer Institut, and which provides a low effort route to neutron scattering experiments that combine electric field with low temperature and magnetic field. The stick allows for voltages up to 5 kV and can be easily adapted for different scattering geometries. We discuss the design consideration and thermal behavior of the stick, and give one example to showcase the abilities of the device. (C) 2014 AIP Publishing LLC.


Publié dans:
Review of Scientific Instruments, 85, 2, 026112
Année
2014
Publisher:
Melville, Amer Inst Physics
ISSN:
1089-7623
Laboratoires:




 Notice créée le 2014-02-13, modifiée le 2019-12-05

n/a:
Télécharger le documentPDF
Lien externe:
Télécharger le documentURL
Évaluer ce document:

Rate this document:
1
2
3
 
(Pas encore évalué)