Improving T2 -weighted imaging at high field through the use of kT -points

PURPOSE: At high magnetic field strengths (B0 >/= 3 T), the shorter radiofrequency wavelength produces an inhomogeneous distribution of the transmit magnetic field. This can lead to variable contrast across the brain which is particularly pronounced in T2 -weighted imaging that requires multiple radiofrequency pulses. To obtain T2 -weighted images with uniform contrast throughout the whole brain at 7 T, short (2-3 ms) 3D tailored radiofrequency pulses (kT -points) were integrated into a 3D variable flip angle turbo spin echo sequence. METHODS: The excitation and refocusing "hard" pulses of a variable flip angle turbo spin echo sequence were replaced with kT -point pulses. Spatially resolved extended phase graph simulations and in vivo acquisitions at 7 T, utilizing both single channel and parallel-transmit systems, were used to test different kT -point configurations. RESULTS: Simulations indicated that an extended optimized k-space trajectory ensured a more homogeneous signal throughout images. In vivo experiments showed that high quality T2 -weighted brain images with uniform signal and contrast were obtained at 7 T by using the proposed methodology. CONCLUSION: This work demonstrates that T2 -weighted images devoid of artifacts resulting from B1 + inhomogeneity can be obtained at high field through the optimization of extended kT -point pulses.


Published in:
Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine, 71, 4, 1478–1488
Year:
2014
Publisher:
Hoboken, Wiley-Blackwell
ISSN:
1522-2594
Keywords:
Laboratories:




 Record created 2013-06-25, last modified 2018-01-28

External link:
Download fulltext
Publisher's version
Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)