Microelectrode arrays for electrophysiological monitoring of hippocampal organotypic slice cultures

A three-dimensional platinum (Pt) microelectrode array embedded on a micromachined silicon (S) substrate (porosity of 13%, via hole diameter of 40 μm) has been developed. Electrodes are 35-μm wide and 20-μm high, spaced 200 μm apart and arranged in an elliptic geometry. Integrated within a microperfusion chamber, the devices were used for stimulation and recording experiments of hippocampal slice cultureS over a period of several days. | A three-dimensional platinum (Pt) microelectrode array embedded on a micromachined silicon (Si) substrate (porosity of 13%, via hole diameter of 40 μm) has been developed. Electrodes are 35-μm wide and 20-μm high, spaced 200 μm apart and arranged in an elliptic geometry. Integrated within a microperfusion chamber, the devices were used for stimulation and recording experiments of hippocampal slice cultures over a period of several days.


Published in:
IEEE Transactions on Biomedical Engineering, 44, 11, 1159-1163
Year:
1997
ISSN:
00189294
Note:
178
Laboratories:




 Record created 2009-05-12, last modified 2018-03-17