  4. Analysis of Lipoproteins by Capillary Zone Electrophoresis in Microfluidic Devices: Assey Development and Surfaces Roughness Measurements
 
research article

Analysis of Lipoproteins by Capillary Zone Electrophoresis in Microfluidic Devices: Assey Development and Surfaces Roughness Measurements

Weiller, B.
•
Ceriotti, L.
•
Shibata, T.
Show more
2002
Analytical Chemistry

The development of a new assay for lipoproteins by capillary electrophoresis in fused-silica capillaries and in glass microdevices is described in this paper. The separation of low-density (LDL) and high-density (HDL) lipoproteins by capillary zone electrophoresis is demonstrated in fused-silica capillaries with both UV absorption and laser-induced fluorescence detection. This separation was accomplished using Tricine buffer (pH 9.0) with methylglucamine added as a dynamic coating. With UV detection, LDL eluted as a relatively sharp peak with a migration time of ∼11 min and HDL eluted as a broad peak with a migration time of 12.5 min. Fluorescence detection of lipoproteins stained with NBD-ceramide was used with the same buffer system to give comparable results. Furthermore, fluorescence staining of human serum samples yielded results similar to the fluorescently stained LDL and HDL fractions, showing that this method can be used to quantify lipoproteins in serum samples. The method was also used to detect lipoproteins in glass micro-CE devices. Very similar results were obtained in microdevices although with much faster analysis times, LDL eluted as a sharp peak at ∼25 s and HDL as a broad peak at slightly longer time. In addition, higher resolution was obtained on chips. To our knowledge, these results show the first separation and detection of lipoproteins in a microfluidic device using native serum samples. Atomic force microscopy was used to characterize the rms surface roughness (Rq) of microfluidic channels directly. Devices with different surface roughness values were fabricated using two different etchants for Pyrex wafers with a polysilicon masking layer. Using 49% HF, the measured roughness is Rq q = 10.9 ± 1.6 nm and with buffered HF (NH4F + HF) the roughness is Rq = 2.4 ± 0.7 nm. At this level of surface roughness, there is no observable effect on the performance of the devices for this lipoprotein separation.

  • Details
  • Metrics
Type
research article
DOI
10.1021/ac011096y
Author(s)
Weiller, B.
Ceriotti, L.
Shibata, T.
Rein, D.
Roberts, M. A.
Lichtenberg, J.
German, J. B.
de Rooij, N. F.  
Verpoorte, E.
Date Issued

2002

Published in
Analytical Chemistry
Volume

74

Issue

7

Start page

1702

End page

1711

Note

278

Editorial or Peer reviewed

REVIEWED

Written at

OTHER

EPFL units
SAMLAB  
Available on Infoscience
May 12, 2009
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/40017
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés