Loading...
research article
Simple dose rate measurements for a very high synchrotron X-ray flux
Dose measurements based on methylene blue (MB) bleaching, widely used for ultraviolet light, can also be applied to X-rays including very high flux levels. This method has been tested by using both MB bleaching and Fricke dosimetry for a conventional monochromatic X-ray source and then for 'white-beam' synchrotron radiation. The results show that MB bleaching dosimetry can easily measure X-ray doses up to at least 10(5) Gy s(-1), as long as the MB concentration is sufficiently high. This condition can be verified from the deviations from linearity of the bleaching versus exposure time.
Type
research article
Web of Science ID
WOS:000265220600013
Authors
Publication date
2009
Published in
Volume
16
Start page
395
End page
397
Peer reviewed
REVIEWED
Available on Infoscience
November 30, 2010
Use this identifier to reference this record