Loading...
research article
Preconcentration Modeling for the Optimization of a Micro Gas Preconcentrator Applied to Environmental Monitoring
2015
This paper presents the optimization of a micro gas preconcentrator (mu-GP) system applied to atmospheric pollution monitoring, with the help of a complete modeling of the preconcentration cycle. Two different approaches based on kinetic equations are used to illustrate the behavior of the micro gas preconcentrator for given experimental conditions. The need for high adsorption flow and heating rate and for low desorption flow and detection volume is demonstrated in this paper. Preliminary to this optimization, the preconcentration factor is discussed and a definition is proposed.
Type
research article
Web of Science ID
WOS:000353429200058
Authors
Camara, Malick
•
Breuil, Philippe
•
•
Viricelle, Jean-Paul
•
Pijolat, Christophe
•
Publication date
2015
Publisher
Published in
Volume
87
Issue
8
Start page
4455
End page
4463
Peer reviewed
REVIEWED
EPFL units
Available on Infoscience
May 29, 2015
Use this identifier to reference this record