Loading...
1993
Human granulocytes detected with a piezoimmunosensor
research article
We have developed a reusable piezoelectric immunosensor for human granulocytes. Three different methods for immobilization of the anti-CD 18 antibody on the gold electrode of the immunosensor were tested. The methods showed some differences in terms of stability and sensitivity. Applying an anti-CD 18 antibody layer onto a 10-MHz AT-cut crystal resulted for 2 × 103 to 3 × 105 cells on the surface in a linear frequency change. The modified crystal was stable for eight weeks when stored desiccated at room temperature.
Type
research article
Author(s)
Date Issued
1993
Published in
Volume
26
Issue
11
Start page
2313
End page
2328
Peer reviewed
REVIEWED
Written at
EPFL
EPFL units
Available on Infoscience
February 21, 2006
Use this identifier to reference this record