Loading...
research article
SECM photography
A novel strategy for imaging commercial black-and-white negative photo films by scanning electrochemical microscopy (SECM) is demonstrated. The imaging is based on using IrCl3 6 as the redox mediator to dissolve silver grains. The SECM probe is biased at a sufficient constant anodic potential (0.8 V vs. Ag QRE) to oxidize IrCl3 6 under diffusion control. SECM images of a patterned band array (2.8 mm 3 mm) and of an ‘‘EPFL” logo (8.6 mm 7 mm) have been recorded with a high resolution from the black-and-white negative photo films. The factors relevant to the quality of images are discussed.
Loading...
Name
sdarticle(2).pdf
Type
Publisher's version
Access type
openaccess
Size
400.11 KB
Format
Adobe PDF
Checksum (MD5)
11695a0e3b7fc1b3de9c855d62a37284