  4. Accuracy of commercial devices and methods for noninvasive estimation of aortic systolic blood pressure a systematic review and meta-analysis of invasive validation studies
 
review article

Accuracy of commercial devices and methods for noninvasive estimation of aortic systolic blood pressure a systematic review and meta-analysis of invasive validation studies

Papaioannou, Theodore G.
•
Karageorgopoulou, Theofani D.
•
Sergentanis, Theodoros N.
Show more
2016
Journal Of Hypertension

Background: Although compelling evidence has established the physiological and clinical relevance of aortic SBP (a-SBP), no consensus exists regarding the validity of the available methods/techniques that noninvasively measure it. Objectives: The systematic review and meta-analysis aimed to determine the accuracy of commercial devices estimating a-SBP noninvasively, which have been validated by invasive measurement of a-SBP. Moreover their optimal mode of application, in terms of calibration, as well as specific technique and arterial site of pulse wave acquisition were further investigated. Methods: The study was performed according to the PRISMA guidelines; 22 eligible studies were included, which validated invasively 11 different commercial devices in 808 study participants. Results: Overall, the error in a-SBP estimation (estimated minus actual value) was -4.49mmHg [ 95% confidence interval (CI): -6.06 to -2.92 mmHg]. The estimated (noninvasive) a-SBP differed from the actual (invasive) value depending on calibration method: by -1.08mmHg (95% CI: -2.81, 0.65 mmHg) and by -5.81mmHg (95% CI: -7.79, -3.84 mmHg), when invasively and noninvasively measured brachial BP values were used respectively; by -1.83 mmHg, (95% CI: -3.32, -0.34 mmHg), and by 7.78mmHg (95% CI: -10.28, -5.28 mmHg), when brachial mean arterial pressure/DBP and SBP/DBP were used, respectively. Conclusion: Automated recording of waveforms, calibrated noninvasively by brachial mean arterial pressure/DBP values seems the most promising approach that can provide relatively more accurate, noninvasive estimation of a-SBP. It is still uncertain whether a specific device can be recommended as 'gold standard'; however, a consensus is currently demanding.

  • Details
  • Metrics
Type
review article
DOI
10.1097/Hjh.0000000000000921
Web of Science ID

WOS:000379448000002

Author(s)
Papaioannou, Theodore G.
Karageorgopoulou, Theofani D.
Sergentanis, Theodoros N.
Protogerou, Athanase D.
Psaltopoulou, Theodora
Sharman, James E.
Weber, Thomas
Blacher, Jacques
Daskalopoulou, Stella S.
Wassertheurer, Siegfried
Show more
Date Issued

2016

Publisher

Lippincott, Williams & Wilkins

Published in
Journal Of Hypertension
Volume

34

Issue

7

Start page

1237

End page

1248

Subjects

accuracy

•

aortic hemodynamics

•

applanation tonometry

•

calibration

•

fluid-filled catheter manometer

Editorial or Peer reviewed

REVIEWED

Written at

EPFL

EPFL units
LHTC  
Available on Infoscience
October 18, 2016
Use this identifier to reference this record
https://infoscience.epfl.ch/handle/20.500.14299/129890
Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés