Parylene C Coated Microelectrodes For Scanning Electrochemical Microscopy
Herein, we present a simple microelectrode preparation methodology consisting in coating a platinum wire or a carbon fiber with a thin insulating Parylene C film (e.g. 1–10 μm), to produce SECM probes with a small and constant probe RG (i.e. ratio between the radius of the insulating sheath and the radius of the active electrode area). After exposition of a fresh active electrode area by blade cutting, a disc shaped electrode is obtained thanks to a protective hot mounting wax layer that avoids Parylene C coating deformation and is easily removed with acetone. Stiffness and straightness of the probe can be tuned by modifying the Parylene C coating thickness and the length of the carbon fiber or platinum wire. This simple electrode preparation method is highly reproducible (c.a. > 90%). The prepared Parylene C coated microelectrodes were characterized by optical microscopy, cyclic voltammetry, scanning electrochemical microscopy (SECM) approach curves and finally applied to SECM imaging of Pt band structures in contact-less and contact mode.
1-s2.0-S0013468613005811-main.pdf
Publisher's version
openaccess
1.69 MB
Adobe PDF
d0fee02dbf587435f5ea2d79700f83e1