Improved calibration of vertical scanning optical profilometers for spherical profiles measurements

Geyl, Roland
•
Schröder, Sven
•
Béguelin, Jeremy
Show more
June 15, 2018
Optical Fabrication, Testing, and Metrology VI

A new method for calibrating optical scanning profilometers is presented. Especially adapted to spherical and aspherical profile measurements, it shows an increase of accuracy bigger than one order of magnitude for radius of curvature measurements. Calibration of vertical scaling is obtained with a reduction of its uncertainty by a factor larger than 2, which also demonstrates the advantage of this method for any surface measurements. Using commercially available reference balls, this method is easily implementable

  • Files
  • Details
  • Metrics
Loading...
Thumbnail Image
Name

Improved calibration of vertical scanning optical profilometers for spherical profiles measurements.pdf

Access type

openaccess

Size

377.17 KB

Format

Adobe PDF

Checksum (MD5)

8f64a2d86e8c0ae52ec57177aaa09919

Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés