A 0.4 nJ Excitation Energy Bridge-to-Digital Converter for Implantable Pulmonary Artery Pressure Monitoring
This paper presents an energy-efficient, duty-cycled, and spinning excitation bridge-to-digital converter (BDC) suitable for measuring the pulmonary artery pressure of heart failure patients with an implantable system. The duty-cycled bridge uses resistances of 6.2 kΩ and, with a supply of 1.2 V, consumes 0.4 nJ excitation energy. A novel spinning method is applied to the bridge and the capacitive DAC simultaneously in such a way to achieve an offset-independent digital output and to eliminate the need for complex instrumentation amplifiers with offset-reduction techniques or calibration. The SAR ADC fabricated in 0.18-µm CMOS consumes 19 nW at 1.2 V. With a sampling rate of 1 kS/ s , the converter achieves the ENOB of 9.2 bits.
2021-10-07
978-1-728172-04-0
REVIEWED
EPFL
Event name | Event place | Event date |
Berlin, Germany | October 7-9, 2021 | |