  4. Automatic, localized in vivo adjustment of all first- and second-order shim coils
 
research article

Automatic, localized in vivo adjustment of all first- and second-order shim coils

Gruetter, R.  
1993
Magnetic Resonance in Medicine

The implementation of a "fast, automatic shimming technique by mapping along projections" (FASTMAP) on a 2.1-Tesla whole-body system is described. The method provides a localized adjustment of all first- (X, Y, Z) and second-order (Z2, ZX, ZY, X2-Y2, 2XY) coils in 2 min. The time savings are achieved by mapping the magnetic field along six projections rather than from whole-imaging data sets. An analysis of noise error propagation suggests that in 64-ml volumes the residual linewidths obtained with the method are negligible when the signal-to-rms-noise ratio is above 30. The initial application of the method to localized 13C, 31P, and 1H spectroscopy of the human brain resulted in linewidths of approximately 2 Hz for 13C (144-ml volume), approximately 2.5 Hz for 31P (36-ml volume) and approximately 4 Hz for 1H (36-ml volume) with symmetric and reproducible lineshapes.

  • Files
  • Details
  • Metrics
Loading...
Thumbnail Image
Name

FASTMAP.PDF

Type

Publisher's Version

Version

Published version

Access type

openaccess

Size

617.57 KB

Format

Adobe PDF

Checksum (MD5)

f240c1b96755a5e1e70780c8133450a1

Logo EPFL, École polytechnique fédérale de Lausanne
  • Contact
  • infoscience@epfl.ch

  • Follow us on Facebook
  • Follow us on Instagram
  • Follow us on LinkedIn
  • Follow us on X
  • Follow us on Youtube
AccessibilityLegal noticePrivacy policyCookie settingsEnd User AgreementGet helpFeedback

Infoscience is a service managed and provided by the Library and IT Services of EPFL. © EPFL, tous droits réservés